Astm D95 Pdf __hot__ [360p]
The condensed liquid falls into a graduated glass trap (often a Dean-Stark trap ). Because water is denser than the solvent and immiscible with it, it sinks to the bottom of the trap, while the excess solvent overflows back into the boiling flask.
The method relies on to separate water from a sample. It is effective for measuring water content ranging from 0% to 25% by volume . For concentrations outside this range, or for specific materials like crude oil, alternative standards like ASTM D4006 are typically used. Astm D95 Pdf
The core of the ASTM D95 method is the . The procedure involves the following steps: The condensed liquid falls into a graduated glass